又來到應用分享時間
這次的問題是Concise Coregel 87H3可以取代Aminex HPX-87H???
發文者: Sary小編
問題 |
某客戶找到用Bio-Rad Aminex HPX-87H column 分析衣康酸的以下條件,請問Concise Coregel 87H3可以取代Aminex HPX-87H ? |
使用條件 |
Mobile phase: 0.5mM H2SO4 Flow rate: 0.6 ml/min Temp: 50 C |
回應 |
Concise原廠回應:我們的Coregel 87H3略好於Aminex HPX-87H,因此如果客戶已經在HPX-87H上進行了這種類型的分析,那麼我們的87H3應該也能正常分析。
如果有更進一步關於樣品混合物(衣康酸和其他成分)的一些細節,我們可能會建議一個更好的替代品。 |
Question |
Customer want to analyze Itaconic acid and provide some information that using Aminex HPX-87H column with below conditions. Could Coregel 87H3 replace Aminex HPX-87H for this analysis? or, do you have orther suggestion? |
Condition |
Mobile phase: 0.5mM H2SO4 Flow rate: 0.6 ml/min Temp: 50 C (she hope the temperature would not be too high) |
Answer |
Our 87H3 is slightly better than the Aminex HPX-87H, so if the customer already performed this type of analysis on the HPX-87H then our 87H3 should work as well (see attached c-gram).
If we have some details about the sample mixture (Itaconic acid and what other components), we might be able to suggest a better alternative. |
參考圖譜
型號 |
品名 |
125-0140 |
Aminex HPX-87H 分離管 7.8x300mm |
ICE-99-9861 |
IC Sep Coregel 87H3 分離管 7.8x300mm |
-想看更多文章
-如果有任何關於應用分析的問題歡迎在下留言喔!
-如果需要詢價請來電02-2776-6931詢問喔!
google-site-verification: googlef4a77e21e4045108.html